Technical ReportFC-TR-2026-01Validation

AVAL: A Validation Pipeline for AVA Mobility

ForesightCares, Inc.Charlotte, NCPublished

13 pages · 12 figures ·4 tables · PDF 2.1 MB

The question · Report 1 of 3 in the AVA Evidence Program

“Is AVA accurate in real community settings?”

Key findings

  • 6.0%

    TUG timing error (MAPE)

    Mean absolute percentage error of AVA’s Timed Up and Go times relative to the mean of two physical-therapist timings, across 62 assessments.

    Source: Abstract; Table 2, p. 8

  • 0.71 s

    Mean absolute TUG error

    Average size of AVA’s TUG timing error against the PT reference; by comparison, the MR4 overestimated TUG time by 1.59 s on average.

    Source: Abstract; Table 2, p. 8

  • 98.5%

    Chair-stand counts matched exactly

    AVA’s count matched the PT count for 64 of 65 participants, and all 65 were within one repetition.

    Source: Section 7.2, p. 10

  • 88.1%

    TUG mobility-category agreement

    Share of 59 participants AVA placed in the same low, fair, or high category as the PTs, versus 71.2% for the MR4 on the same participants.

    Source: Abstract; Table 3, p. 8

Abstract

Mobility tests such as the 30-second chair stand and the Timed Up and Go (TUG) are widely used to identify older adults at elevated risk of falling. Both tests are simple, but they depend on a trained observer to count repetitions or run a stopwatch, which limits how often and how widely they can be given. AVA Mobility is a vision-based assessment system that automates these tests using a single iPad camera. It performs pose estimation and movement-event detection entirely on the device, with no wearable sensors, external hardware, or cloud processing. AVA performs well in controlled testing, but its accuracy in the community settings where it is meant to be used had not been formally measured.

This report introduces AVAL, a validation pipeline that compares AVA’s results against measurements made by physical therapists (PTs) and against a commercial wearable motion-capture system (Noraxon MR4). Data were collected from 66 older adults across four community sites between June and October 2025. AVAL links each participant’s video to their reference measurements, runs AVA on the video, reconstructs the test outcome, and computes error metrics. TUG timing is evaluated with MAPE-based accuracy, mean absolute error, root mean squared error, and 95th-percentile error. Chair-stand counts are evaluated by the share of predictions that match the clinical count exactly or within one repetition. We also compare the low, fair, and high mobility categories each system assigns using CDC STEADI thresholds.

On the TUG, AVA’s timings had a mean absolute percentage error (MAPE) of 6.0% relative to PT timing, with a mean absolute error of 0.71 s; the MR4 overestimated TUG time by 1.59 s on average. On the chair stand, AVA matched the clinical count exactly for all but one participant, and every prediction was within one repetition. AVA also agreed more closely than the MR4 with PT-assigned mobility categories on both tests (TUG: 88.1% vs. 71.2%; chair stand: 100% vs. 87.5%, on the same participants). These results indicate that AVA can deliver accurate, scalable, and privacy-preserving mobility assessment in real-world community settings.

Keywords: fall risk · mobility assessment · pose estimation · Timed Up and Go · 30-second chair stand · STEADI · validation

Why it matters

The Timed Up and Go and the 30-second chair stand are short, widely used tests for identifying older adults at elevated risk of falling. Both normally need a trained observer to count repetitions or run a stopwatch. This report tested whether AVA Mobility, which runs these tests with a single iPad camera, matches physical therapists’ measurements in everyday community spaces rather than a lab. Data came from 66 older adults at four community sites, and the same tests were also recorded with a commercial wearable motion-capture system (the Noraxon MR4) for comparison.

AVA’s Timed Up and Go times differed from the therapists’ times by 0.71 seconds on average, with errors centered close to zero, and the pattern was similar at all four sites. On the chair stand, AVA’s count matched the therapists’ count for 64 of 65 participants, and the one remaining count was off by a single repetition. When results were sorted into low, fair, and high mobility categories, AVA agreed with the therapists for 88.1% of Timed Up and Go results and 100% of chair-stand results, compared with 71.2% and 87.5% for the motion-capture system, which tended to report longer times.

Clinicians usually act on the mobility category rather than the raw time or count, so category agreement is the result most relevant to screening decisions. The motion-capture system needs sixteen body-worn sensors, calibration, and a trained operator, roughly 30 to 40 minutes of setup per participant; AVA ran the same assessments on one iPad in under five minutes, with processing done on the device. The study measured agreement with therapists, not fall outcomes, and its sample was moderate in size and drawn from one region.

Explore the results

Interactive views of numbers published in the report. Hover or focus a mark for details.

Agreement with physical-therapist mobility categories

Percentage of participants AVA and the MR4 placed in the same low, fair, or high mobility category as the physical therapists, on the same participants. Chair-stand categories follow CDC STEADI normative values; TUG categories (≤11 s high, ≥15 s low) were defined for this study.

  • AVA
  • Noraxon MR4
Timed Up and Gon = 59; AVA 52/59, MR4 42/59; Cohen’s κ 0.79 vs. 0.55
30-second chair standn = 56; AVA 56/56, MR4 49/56; Cohen’s κ 1.00 vs. 0.78
Source: Table 3, p. 8
View data table
AssessmentAVANoraxon MR4Notes
Timed Up and Go88.1%71.2%n = 59; AVA 52/59, MR4 42/59; Cohen’s κ 0.79 vs. 0.55
30-second chair stand100%87.5%n = 56; AVA 56/56, MR4 49/56; Cohen’s κ 1.00 vs. 0.78

Selected figures

From the report. Select a figure to enlarge it, or jump to its page in the full report.

Figure 5. The AVAL pipeline. Each recorded video and its metadata are processed by AVA Mobility to produce AVA’s assessment. The same assessment is scored by a physical therapist and recorded by the MR4 motion-capture system. AVAL compares the three results for each participant and summarizes the errors in the analysis stage. p. 6 in the report →
Figure 6a. Predicted versus manually measured TUG times for AVA. Tolerance bands mark ±5% and ±10% of the reference value, and error bars show the range across observers. p. 9 in the report →
Figure 6b. Predicted versus manually measured TUG times for the MR4 motion-capture system. Tolerance bands mark ±5% and ±10% of the reference value, and error bars show the range across observers. p. 9 in the report →
Figure 8. Signed TUG errors (system minus PT mean) by community for AVA and the MR4. p. 9 in the report →
Figure 12b. Mobility-category agreement on the Timed Up and Go between each automated system (MR4 left, AVA right) and the manual measurements. Rows show the category from manual measurement, and columns show the category each system assigned. p. 12 in the report →

Full report

The PDF is the version of record. Search, follow internal links, or jump to any page.

Limitations

  • Very dim or very bright lighting, people or objects passing between the camera and the participant, incorrect iPad positioning, and clothing that blends into the background can affect measurement quality.
  • Some recordings were excluded because of calibration failures or missing data in the reference measurements.
  • The two PT observers’ start and stop cues and their blinding to AVA and MR4 outputs were not documented, and AVA and the PTs applied different rules to the final chair-stand repetition.
  • The sample of 66 participants is moderate in size, weighted toward women (48 of 66), and drawn from four sites in one region.
  • The 4-stage balance test was recorded but not analyzed.

Ethics, funding & disclosures

The study was approved by the Advarra Institutional Review Board (protocol Pro00080873) and conducted in accordance with the Declaration of Helsinki. All participants gave informed consent before taking part, including consent to video recording and to wearing the MR4 sensors.

This report was prepared by ForesightCares, Inc., which develops AVA.

Responsible use

AVA supports mobility awareness and screening. It does not diagnose conditions, replace clinical evaluation, or serve as a standalone clinical decision tool. Results describe the study population and settings reported here.